A CE-marked fracture detection model, retrained after launch
A CE-marked AI that flags fractures gets better after retraining. Whether it can ship depends on what the manufacturer planned before the first certificate.
Last reviewed: 28 October 2026 · First published on LinkedIn: 26 October 2026
01
A startup sells software that flags fractures on X-rays. It is CE marked as a class IIa medical device and used in a dozen hospitals.
A year later the team has ten times more images. They retrain the model, and sensitivity rises by four points: fewer missed fractures. Everyone wants to ship it.
Then someone asks the right question: is this still the device that was certified?
02
03
Four kinds of change to the same device. What decides is how much the change affects performance, and whether it was planned.
| Variant | What changes | MDR | EU AI Act | From |
|---|---|---|---|---|
| 1 | A user-interface fix; the model is unchanged | Usually not significant; assessed and documented under change control | Not substantial | MDR: now · AI Act: 2 Aug 2028 |
| 2 | The model is retrained on new data; performance changes; nothing was planned | Likely significant: the notified body hears about it before release | Substantial modification: new conformity assessment | MDR: now · AI Act: 2 Aug 2028 |
| 3 | The same retraining, inside an envelope defined at the first assessment: data sources, frequency, performance limits, validation protocol | Still agreed with the notified body, but reviewed against an existing plan | Not substantial, as a pre-determined change | MDR: now · AI Act: 2 Aug 2028 |
| 4 | The model keeps learning in the hospitals, without a release | Hard to certify today; notified bodies generally expect locked models | Acceptable only for changes pre-determined at the initial assessment Art. 43(4) | MDR: now · AI Act: 2 Aug 2028 |
Changing the indication, for example adding children or a new body region, is a change of intended purpose. No envelope covers it: it needs a new assessment under both regulations.
04
The startup split the release. The interface fixes shipped as a minor version under its normal change control. The retrained model waited.
In the meantime the team wrote the envelope into the technical documentation: which data sources may be added, how often the model may be retrained, the performance limits a new version must meet on a fixed test set, and the validation that must pass before release. They agreed the approach with their notified body, and prepared to carry it into the AI Act documentation for 2028.
The next retraining will still be checked. But it will be checked against a plan, not discovered as a surprise.
05
06
Questions
If your product sits close to one of these lines, the first question is what exactly it does. Write to me if you want to talk it through.
Case files are study material, not legal advice. Cases are composite or anonymised; any resemblance to a specific company is not intended.