Case files · #4

Better is still different

A CE-marked fracture detection model, retrained after launch

A CE-marked AI that flags fractures gets better after retraining. Whether it can ship depends on what the manufacturer planned before the first certificate.

Last reviewed: 28 October 2026 · First published on LinkedIn: 26 October 2026

Composite caseMedical devicesArt. 3(23)Art. 43(4)Change control

01

The scenario

A startup sells software that flags fractures on X-rays. It is CE marked as a class IIa medical device and used in a dozen hospitals.

A year later the team has ten times more images. They retrain the model, and sensitivity rises by four points: fewer missed fractures. Everyone wants to ship it.

Then someone asks the right question: is this still the device that was certified?

02

Roles and classification

  • The startup is the manufacturer under the Medical Device Regulation and the provider under the AI Act Art. 3(3). The hospitals using it professionally are deployers Art. 26.
  • MDR: the notified body must be informed of planned changes that could affect the safety or performance of the device or its approved design, and assesses them before they are implemented MDR Annex IX. A retrained model changes performance, even when it improves it.
  • AI Act: the device is high-risk because it needs a notified body Art. 6(1), Annex I, from 2 August 2028. A substantial modification is a change not foreseen or planned in the initial conformity assessment that affects compliance or the intended purpose Art. 3(23), and it requires a new conformity assessment Art. 43(4).
  • The planned-change route: changes pre-determined by the provider at the initial assessment and described in the technical documentation are not substantial modifications Arts. 3(23), 43(4). For Annex I products the assessment follows the MDR procedure, with the AI Act requirements included Art. 43(3).

03

The variants

Four kinds of change to the same device. What decides is how much the change affects performance, and whether it was planned.

VariantWhat changesMDREU AI ActFrom
1A user-interface fix; the model is unchangedUsually not significant; assessed and documented under change controlNot substantialMDR: now · AI Act: 2 Aug 2028
2The model is retrained on new data; performance changes; nothing was plannedLikely significant: the notified body hears about it before releaseSubstantial modification: new conformity assessmentMDR: now · AI Act: 2 Aug 2028
3The same retraining, inside an envelope defined at the first assessment: data sources, frequency, performance limits, validation protocolStill agreed with the notified body, but reviewed against an existing planNot substantial, as a pre-determined changeMDR: now · AI Act: 2 Aug 2028
4The model keeps learning in the hospitals, without a releaseHard to certify today; notified bodies generally expect locked modelsAcceptable only for changes pre-determined at the initial assessment Art. 43(4)MDR: now · AI Act: 2 Aug 2028

Changing the indication, for example adding children or a new body region, is a change of intended purpose. No envelope covers it: it needs a new assessment under both regulations.

04

What the company did

The startup split the release. The interface fixes shipped as a minor version under its normal change control. The retrained model waited.

In the meantime the team wrote the envelope into the technical documentation: which data sources may be added, how often the model may be retrained, the performance limits a new version must meet on a fixed test set, and the validation that must pass before release. They agreed the approach with their notified body, and prepared to carry it into the AI Act documentation for 2028.

The next retraining will still be checked. But it will be checked against a plan, not discovered as a surprise.

05

Lessons for teams

  1. Better performance is still different performance. Treat every retrained model as a design change until your change control says otherwise.
  2. Define the envelope before the first certificate: data, frequency, performance limits, validation. Under the AI Act, planned changes are not substantial modifications.
  3. Keep a fixed, versioned test set that never enters training. It is what proves a new version stays inside the envelope.
  4. Separate model releases from software releases, so that a user-interface fix never waits for a model assessment, and the reverse.
  5. A change of indication is never inside an envelope. Plan it as a new assessment.

06

Open questions

  • The MDR has no formal equivalent of a pre-determined change plan. Will MDCG guidance align it with the AI Act before 2028, or will notified bodies keep handling it case by case?
  • How wide can an envelope be before a notified body considers it a blank cheque?
  • The FDA already accepts predetermined change control plans for AI-enabled devices. Will manufacturers selling in both markets converge on a single plan?

Related

Questions

A case like yours?

If your product sits close to one of these lines, the first question is what exactly it does. Write to me if you want to talk it through.

Case files are study material, not legal advice. Cases are composite or anonymised; any resemblance to a specific company is not intended.